RC-900B Ventilator
March 18, 2020
Show all

S1100A
ICU Ventilator

The ICU ventilator S1100A is a respiratory support machine which is used within the ICU department and for emergency treatment. As well as temporarily supporting the functionality of the human lungs, this respirator features nebulization, which administers medication in the form of a mist for patients to then inhale into the lungs.

This system includes a range of configurations including the likes of multiple ventilation modes to support the patients needs, all controlled via the large 15” TFT touch screen display. For more information on the product, follow product details.

Manufacturer: Perlove Medical

  • 15” TFT touch screen displays the ventilation parameters, alarm information, and oscillograms.
  • Multiple ventilation modes, suitable for ICU, emergency department and operating theatre etc: (IPPV, A/C, PCV, SIMV, PSV, SPONT/CPAP, SIGH, MANUAL)
  • 5 oscillograms for your choice, 3 of them can be displayed on the screen in the same time.
  • ETCO2 is the standard configuration
  • The humidifier can heat and wet breathing gas
  • Rapid oxygen supply, automatically offer high flow rate oxygen within two minutes
  • Nebulization make medicine into small liquid and particle, easier and quicker for the patient to breathe
  • Reuseable high-temperature resistance breathing circuit
Ventilation mode
IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV, SIGH, MANUAL
Ventilator parameter
Tidal volume (Vt) 0-200ml
Frequency (Freq) 1min – 100min
Oxygen concentration 21% – 100%
I:E 4:1 – 1:8
PEEP 0cmH2O – 40cmH2O
Pressure triggering sensitivity (Ptr) -20cmH2O – 20cmH2O (Based on PEEP)
Flow trigger sensitivity (Ftr) 0.5 L/min – 30 L/min
Pressure control (PC) 5cmH2O – 80cmH2O
Pressure support (PS) 0cmH2O – 80cmH2O
SIGH 0 (off) 1/100 5/100
Apnea ventilation OFF, 5s – 60s
Pressure limit 20cmH2O – 100cmH2O
Monitoring parameter
Frequency (Freq) 0/min – 100/min
Tidal volume (Vt) 0mL – 2500mL
MV 0 L/min – 99 L/min
Airway pressure 0 cmH2O – 100 cmH2O
Dynamic lung compliance testing 1 mL/cmH2O – 1000 mL/chH2O
ETCO2 concentration 0mm Hg – 152 mmHg (0% – 20%)
Oxygen concentration 15% – 100%
Dimensions
Main body
Size L 560 W560 H 605 mm
Weight G.W. 40KG, N.W 22KG
Air compressor
Size L 683 W 678 H1140mm
Weight G.W. 105KG, N.W. 70KG
Alarm and protection
AC power failure alarm Power failure or no connection
Internal backup battery low voltage alarm ≤11.3 ±  0.3V
No tidal volume No tidal volume within 6 second
High minute volume alarm 5 L/min – 99L/min
Low minute volume alarm 1 L/min – 30L/min
High airway pressure alarm 20 cmH2O – 100cmH2O
Low airway pressure alarm 0 cm H2O – 20cmH2O
High oxygen concentration alarm 19% -100%
Low oxygen concentration alarm 18% -99%
Continuous pressure alarm 5-60 second
Fan error Show on screen
Oxygen deficit Show on screen
Maximum limited pressure <21.5 kPa
Working condition
Gas source O2, Air
Pressure 280 kPa – 600kPa
Voltage 220V ± 22V
Power frequency 50 Hz ± 1 Hz
Input power

900 VA (With air compressor)

250 VA (Without air compressor)

Oscillogram
P-T Pressure-time
F-T Flow-time
V-T Volume-time
ETCO2-T End-tidal CO2-time
P-V Loop Pressure-volume Loop
F-V Loop Flow-volume Loop
P-F Loop Pressure-flow Loop